Charles Darwin University

CDU eSpace
Institutional Repository

 
CDU Staff and Student only
 

Point-of-care testing of capillary glucose in the exclusion and diagnosis of diabetes in remote Australia

Marley, Julia V., Davis, Stephanie, Coleman, Kerryn, Hayhow, Bradleigh D., Brennan, Greg, Mein, Jacki K., Nelson, Carmel, Atkinson, David and Maguire, Graeme P. (2007). Point-of-care testing of capillary glucose in the exclusion and diagnosis of diabetes in remote Australia. Medical Journal of Australia,186(10):500-503.

Document type: Journal Article
Citation counts: Google Scholar Search Google Scholar
Attached Files (Some files may be inaccessible until you login with your CDU eSpace credentials)
Name Description MIMEType Size Downloads
Download this reading Maquire_2576.pdf Published version application/pdf 266.83KB 124
Reading the attached file works best in Firefox, Chrome and IE 9 or later.

Title Point-of-care testing of capillary glucose in the exclusion and diagnosis of diabetes in remote Australia
Author Marley, Julia V.
Davis, Stephanie
Coleman, Kerryn
Hayhow, Bradleigh D.
Brennan, Greg
Mein, Jacki K.
Nelson, Carmel
Atkinson, David
Maguire, Graeme P.
Journal Name Medical Journal of Australia
Publication Date 2007
Volume Number 186
Issue Number 10
ISSN 0025-729X   (check CDU catalogue open catalogue search in new window)
Start Page 500
End Page 503
Place of Publication Australia
Publisher Australasian Medical Publishing Company Pty. Ltd.
HERDC Category C1 - Journal Article (DEST)
Abstract Objectives: To determine the utility of point-of-care (POC) capillary blood glucose measurements in the diagnosis and exclusion of diabetes in usual practice in primary health care in remote areas.

Design:
Cross-sectional study comparing POC capillary glucose results with corresponding venous glucose levels measured in a reference laboratory.

Participants:
200 participants aged 16–65 years enrolled: 198 had POC capillary glucose measurements; 164 also had acceptable venous glucose laboratory results.

Setting:
Seven health care sites in the Kimberley region of Western Australia from May to November 2006.

Main outcome measures:
Concordance and mean differences between POC capillary blood glucose measurement and laboratory measurement of venous blood glucose level; POC capillary blood glucose equivalence values for excluding and diagnosing diabetes, and their sensitivity, specificity and positive-predictive value

Results:
The concordance between POC and laboratory results was high (ρ = 0.93, P < 0.001). The mean difference in results was 0.48 mmol/L (95% CI, 0.23–0.73; limits of agreement, − 2.6 to 3.6 mmol/L). The POC capillary glucose equivalence values for excluding and diagnosing diabetes were < 5.5 mmol/L (sensitivity, 53.3%; specificity, 94.4%; positive-predictive value, 88.9%; for a venous value of < 5.5 mmol/L) and ≥ 12.2 mmol/L (sensitivity, 83.3%; specificity, 99.3%; positive-predictive value, 95.2%; for a venous value of ≥ 11.1 mmol/L), respectively. While the choice of glucometer and whether or not patients were fasting altered these results, they did not have a significant influence on the diagnostic utility of POC glucose measurement in this setting.

Conclusion:
POC capillary blood glucose analysers can be used as part of the process of diagnosing and excluding diabetes in remote rural communities using these locally established capillary equivalence values.

Keywords Blood-glucose
complications
population
Mellitus
Description for Link Link to published version
URL https://www.mja.com.au/journal/2007/186/10/point-care-testing-capillary-glucose-exclusion-and-diagnosis-diabetes-remote#authors


© copyright

Every reasonable effort has been made to ensure that permission has been obtained for items included in CDU eSpace. If you believe that your rights have been infringed by this repository, please contact digitisation@cdu.edu.au.

 
Versions
Version Filter Type
Access Statistics: 151 Abstract Views, 125 File Downloads  -  Detailed Statistics
Created: Fri, 12 Sep 2008, 08:35:25 CST by Administrator